Repeated blood glucose measures using a novel portable glucose meter
Physiology & Behavior
|April 1, 1995
Summary
A new portable blood glucose meter accurately measures mouse blood glucose levels, enabling rapid, repetitive sampling for the mouse insulin tolerance test (ITT). This device offers high precision and reliability for preclinical research.
Area of Science:
- Preclinical research
- Animal models
- Diabetes research
Background:
- Accurate blood glucose monitoring is crucial for preclinical diabetes research.
- Existing methods for mouse blood glucose measurement can be time-consuming or invasive.
- Adapting human physiological tests to animal models requires reliable monitoring tools.
Purpose of the Study:
- To evaluate a novel portable blood glucose meter for repetitive and rapid glucose measurements in mice.
- To validate the meter's performance against a conventional glucose analyzer.
- To assess the feasibility of using this device for a mouse insulin tolerance test (ITT).
Main Methods:
- A novel portable blood glucose meter was used for serial blood glucose measurements in mice.
- Comparative analysis was performed against a standard automated glucose analyzer.
- Measurements were taken from capillary and trunk whole blood.
- The mouse insulin tolerance test (ITT) protocol involved repeated sampling post-insulin injection.
Main Results:
- The portable meter demonstrated high correlation with the conventional analyzer for both capillary and trunk blood.
- Measurement variability over five consecutive determinations was approximately 3%.
- Repetitive measurements could be obtained at intervals of about 70 seconds.
Conclusions:
- The novel portable blood glucose meter is precise and reliable for measuring mouse blood glucose.
- This device facilitates rapid, repetitive sampling, making it suitable for tests like the mouse ITT.
- The meter offers a valuable tool for advancing preclinical diabetes research in rodent models.