Type testing of the Siemens Plessey electronic personal dosemeter
1Health Physics Department of Ontario Hydro, Whiby, Canada.
This study assessed the Electronic Personal Dosemeter (EPD) for radiation monitoring. While generally acceptable, the EPD showed deficiencies in response time, electromagnetic field sensitivity, durability, and alarm volume.
Area of Science:
- Health Physics
- Radiation Detection and Measurement
Background:
- Personal dosimeters are crucial for monitoring radiation exposure.
- New electronic personal dosemeters (EPDs) require rigorous performance evaluation.
Purpose of the Study:
- To conduct a laboratory assessment of a new Electronic Personal Dosemeter (EPD) from Siemens Plessey Controls Limited.
- To evaluate the EPD's radiological and environmental performance against industry standards.
Main Methods:
- Tested radiological performance (accuracy, linearity, response times, etc.).
- Assessed environmental factors (temperature, humidity, electromagnetic fields, impacts).
- Compared results against Ontario Hydro and International Electrotechnical Commission (IEC) standards.
Main Results:
- The EPD generally met most standard requirements.
- Identified deficiencies: slow response time, high-frequency electromagnetic field sensitivity, poor drop resistance, and inadequate alarm volume.
- Observed potential under-response to low-energy beta rays and reliability issues with pre-production units.
Conclusions:
- The EPD shows promise but requires improvements for widespread adoption.
- Specific areas for enhancement include response speed, electromagnetic compatibility, physical robustness, and audibility.
- Further validation is needed, especially concerning low-energy beta response and pre-production unit reliability.
More Related Videos
07:44Characterization of Full Set Material Constants and Their Temperature Dependence for Piezoelectric Materials Using Resonant Ultrasound Spectroscopy
Published on: April 27, 2016
07:43A Simple Approach to Perform TEER Measurements Using a Self-Made Volt-Amperemeter with Programmable Output Frequency
Published on: October 5, 2019
Related Concept Videos
Ammeter
When an ammeter is used to measure the...
Voltmeter
An ideal voltmeter would have infinite resistance, so connecting it between two points in a circuit would not alter any of the currents. Real voltmeters always have finite...
Potentiometer
Suppose the emf of a battery needs to be measured. If the battery is directly connected to a standard voltmeter, the measured quantity is...
Wheatstone Bridge
Thus, for accurate resistance measurements, a...
Electrical Transport
Electronic Distance Measuring Instruments