Related Experiment Videos
Vibration in metal and non-metal incubators
J M Youngblut1, W Lewandowski, G R Casper
1Frances Payne Bolton School of Nursing, Case Western Reserve University, Cleveland, Ohio 44106-4904.
Abstract:
The purpose of this study was to determine the amount of vibration transmitted to the surface of an incubator mattress. Empty incubators with metal (n = 12) and non-metal (n = 12) bases were monitored for vibration levels when the incubators were turned "off" and when they were turned "on." High levels of low-frequency vibration were detected in both types of incubators in both conditions. The metal incubators transmitted significantly less vibration to the mattress than did the non-metal incubators at several frequencies in the "off," the "on," and the "adjusted" conditions. These results suggest that infants experience significant whole-body vibration while lying in incubators.