Clinical use of a concomitant boost technique using a gypsum compensator
K J Weeks1, V R Arora, K A Leopold
1Department of Radiation Oncology, Duke University Medical Center, Durham, NC 27710.
Purpose:
To develop a clinical procedure to treat field within a field (concomitant boost) portals with a single compensated field.
Methods And Materials:
An ordinary manual cerrobend block former was used to produce styrofoam molds from simulator film data. A special gypsum compound was poured into the molds. The compensator block is independently mounted to the treatment machine via a custom-made compensator holder.
Results:
Measurements confirm that the inhomogeneous dose distribution has been reliably delivered via this technique. The accuracy of placement of the high dose region is sufficient for clinical use.
Conclusion:
The procedure enables the concomitant boost effect to be easily implemented in the clinic without increasing clinical setup time.