Related Experiment Videos
[Transferability of results in hematologic determinations]
V Alvarez1, A Hernández, C V Jiménez
1Servicio Análisis Clínicos, C.A.P. Cornellá, Bellaterra, Barcelona.
Summary
Laboratory results for common blood counts show transferable inaccuracy between facilities, but not all constituents are consistently reliable. Patient blood samples can help assess some inaccuracies when control materials are unavailable.
Area of Science:
- Clinical Laboratory Science
- Hematology
- Quality Assurance in Healthcare
Background:
- Inter-laboratory quality control is crucial for ensuring the reliability and transferability of hematological test results.
- Variations in analytical methods and equipment can lead to discrepancies in laboratory findings.
Purpose of the Study:
- To evaluate the inaccuracy of hematological test results across multiple Servei Catalá de la Salut laboratories.
- To determine if quality objectives are met and if results are transferable between these laboratories.
Main Methods:
- Assessed inaccuracy using inter-laboratory quality control programs for two types of hematological analyzers (volumetric and light dispersion).
- Conducted inter-series and intra-series inaccuracy studies using commercial control materials and fresh patient blood.
- Calculated monthly coefficients of variation and compared analytical inaccuracy in patient samples versus control materials.
Main Results:
- Inaccuracy quality limits were exceeded for several hematological constituents.
- Quality objectives were met for red cell count, white cell count, MCV, granulocytes, and lymphocytes in the inter-series study.
- Fresh patient blood assays can assess intra-series inaccuracy for red and white cell counts, but not platelet counts.
Conclusions:
- Inaccuracy of studied blood constituents is not universally transferable between laboratories.
- Inaccuracy for common hematological values is transferable.
- Fresh patient blood can be used to assess intra-series inaccuracy for red and white cell counts when control material is absent.
- Patient samples are not recommended for controlling white cell differential counts in volumetric systems.