Monte Carlo modelling of neutron depth dose distributions in optimizing prompt gamma in vivo neutron activation
T D Utteridge1, A C Ralston, S A Wallace
1Department of Medical Physics, Royal Adelaide Hospital, Australia.
Abstract:
Optimization of prompt gamma in vivo neutron activation analysis systems is best achieved using Monte Carlo simulation. In this study the modelling of the dimensions and materials for source holders and collimators is described and compared with experimentally derived results where feasible. Results show that valid depth doses are obtained by modelling only the central part of an IVNAA system and that the use of borated paraffin as a reflector provides acceptable thermal fluence and depth dose.