Related Experiment Videos
Application of biotin-streptavidin technology in developing a xanthine biosensor based on a self-assembled
M Rehák1, M Snejdárková, M Otto
1Institute of Animal Biochemistry and Genetics, Slovak Academy of Sciences, Ivanka pri Dunaji.
Biosensors & Bioelectronics
|January 1, 1994
Abstract:
We have designed an amperometric xanthine minibiosensor based on a supported biotinylated phospholipid membrane. To this membrane streptavidin-modified xanthine oxidase was coupled. The fabricated biosensor corresponds in shape and size to a needle of 0.3 mm in diameter. The assay is based on the electrochemical detection of enzymatically generated hydrogen peroxide. The response to xanthine was linear up to 1 mmol.l-1 with a detection limit of 0.02 mmol.l-1. The response time was less than one minute and the biosensor was stable for at least five days.