Metallic spinal artifacts in magnetic resonance imaging
A R Vaccaro1, R M Chesnut, G Scuderi
1Orthopaedic Research Laboratory, University of California, San Diego.
Spine
|June 1, 1994
Summary
Magnetic resonance imaging artifacts from surgical metals vary. Gradient echo imaging shows stainless steel causes less artifact than other metals, making spin echo sequences preferable for visualizing spinal structures post-surgery.
Area of Science:
- Medical Imaging
- Biomaterials Science
- Neurosurgery
Background:
- Surgical implants, including metallic components, can cause artifacts in magnetic resonance imaging (MRI).
- Understanding artifact susceptibility is crucial for accurate postoperative spinal imaging.
- Previous studies have explored metallic artifact generation, but comparative analysis across different sequences and alloys is ongoing.
Purpose of the Study:
- To quantitatively compare void artifact susceptibility of aluminum, titanium, and stainless steel surgical metals using MRI.
- To evaluate artifact differences between spin echo and gradient echo sequences.
- To assess the impact of stainless steel alloy composition on MRI artifact.
Main Methods:
- Human thoracic spine models with deposited metallic filings were imaged using spin echo and gradient echo sequences.
- Void artifact was quantified and compared across different metallic alloys and a liquid paraffin control.
- Two blinded neuroradiologists rated artifact severity on a scale of 1 to 4.
Main Results:
- Artifact magnitude generally correlated with metal magnetism.
- 316L stainless steel produced less artifact than 304 stainless steel on gradient echo imaging due to nickel content stabilizing iron.
- No significant artifact was observed with any metal in all spin echo sequences.
Conclusions:
- Stainless steel should be suspected if significant artifact is seen on postoperative gradient echo imaging.
- Spin echo techniques are recommended for optimal visualization of the spinal cord and soft tissues in the presence of suspected metallic artifacts.
- The choice of MRI sequence is critical for minimizing artifact and ensuring diagnostic accuracy after spinal surgery involving metallic implants.