Beam profile analysis using GafChromic films

M Sanders1, J Sayeg, C Coffey

  • 1Neurosurgery Department, University of Kentucky Medical Center, Lexington.

Stereotactic and Functional Neurosurgery
|January 1, 1993
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

SU-E-T-593: VMAT Vs. IMRT for Treatment of Neoplasms of the Brain: Dosimetric Quality and Delivery.

Medical physics·2017
Same author

SU-E-T-407: Freedom from Conventional: Application of RapidArc for SBRT Lung Treatments.

Medical physics·2017
Same author

SU-E-T-429: Image-Guided Eye Plaque Brachytherapy Optimization: Implications for Patients at 2-Year Follow-up.

Medical physics·2017
Same author

TU-C-218-02: Effective Oncology Physics Education.

Medical physics·2017
Same author

Collaborative meta-analysis finds no evidence of a strong interaction between stress and 5-HTTLPR genotype contributing to the development of depression.

Molecular psychiatry·2017
Same author

Substance use in adulthood following adolescent self-harm: a population-based cohort study.

Acta psychiatrica Scandinavica·2014
Same journal

Stereotactic Accuracy and Operative Metrics Utilizing the ClearPoint Frameless Intraoperative-CT Image-Based Guidance: A Multi-Institution Analysis.

Stereotactic and functional neurosurgery·2026
Same journal

A Fully MRI-Compatible, Non-Electrical System for Near Real-Time Tremor Monitoring During MR-guided Focused Ultrasound.

Stereotactic and functional neurosurgery·2026
Same journal

Metric choice and programming-output concordance in DBS lead reconstruction robustness.

Stereotactic and functional neurosurgery·2026
Same journal

Trends In Deep Brain Stimulation Surgery in The United States.

Stereotactic and functional neurosurgery·2026
Same journal

Successful MRgFUS VIM Thalamotomy for essential tremor nearby cavernous malformation: image case report.

Stereotactic and functional neurosurgery·2026
Same journal

Deep Brain Stimulation for Sleep-Wake Regulation: A Narrative Review.

Stereotactic and functional neurosurgery·2026
See all related articles

GafChromic film dosimetry accurately evaluated Leksell Gamma Knife beam profiles, matching conventional methods. This rapid technique is suitable for quality assurance programs in radiation oncology.

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Radiosurgery

Background:

  • Leksell Gamma Knife acceptance testing requires precise beam profile analysis.
  • Conventional film dosimetry is time-consuming and may involve external analysis.
  • Developing efficient in-house dosimetry methods is crucial for quality assurance.

Purpose of the Study:

  • To evaluate the accuracy and efficiency of GafChromic film dosimetry for Leksell Gamma Knife beam profiles.
  • To compare GafChromic film dosimetry results with conventional film analysis.
  • To assess the suitability of GafChromic films for routine quality assurance.

Main Methods:

  • GafChromic films were irradiated to 100 Gray for each collimator size and plane.
  • Films were analyzed using a scanning helium-neon laser densitometer (5-10 micron aperture).

Related Experiment Videos

  • Digitized density profiles were curve-fitted and converted to relative dose using a Cobalt-60 calibration curve.
  • Main Results:

    • GafChromic film dosimetry results showed excellent agreement with conventional film analysis by Leksell.
    • The method provided accurate data for beam profiles across different collimator sizes.
    • Analysis using GafChromic films was demonstrated to be rapid.

    Conclusions:

    • GafChromic film dosimetry is an accurate and efficient alternative for Leksell Gamma Knife beam profile analysis.
    • This technique can be readily integrated into a comprehensive quality assurance program.
    • The findings support the use of GafChromic films for routine dosimetry in radiosurgery.