Related Experiment Video
Updated: Aug 3, 2026

Synthesis of Non-uniformly Pr-doped SrTiO3 Ceramics and Their Thermoelectric Properties
Published on: August 15, 2015
Comparison of the thermal tissue effects produced by aged sapphire and silica hemispherical tips
D D Royston1, J H Torres, S Thomsen
1Food and Drug Administration, Center for Devices and Radiological Health, Rockville, MD 20857.
Abstract:
This study evaluated the performance of sapphire and fused silica hemispherical tips under the same exposure conditions. Lesions produced in the chicken breast and a blood field were sectioned for light and transmission polarizing microscopy. Lesion size and thermal damage area were recorded as a function of the tips accumulated exposure. The tips transmission was measured after every 1,000 J of exposure. Fused silica tips lasted for approximately 5,000 J and experienced significant surface and transmission deterioration. The sapphire hemispherical tips lasted for > 12,000 J with no surface and transmission deterioration. Lesions produced with the fused silica tips generally increased in depth with use, and depths of 6 mm were common. Lesions produced by the sapphire tips were subsurface spherical areas of coagulation with the tissue surface relatively intact. This difference in resulting lesions may be attributed to the higher thermal conductivity of sapphire.
More Related Videos
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
07:05In Vitro Evaluation of The Effects Of Er,Cr:YSGG and Diode Lasers Used on Titanium Cylinder