Variation in output with a fixed dose setting on a 4-MV linear accelerator
Related Experiment Video
Updated: Aug 9, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 10, 2014
Abstract:
Variations in the output with a fixed monitor unit setting as a function of magnetron power of a 4-MV linear accelerator has been studied. The relationship between radiation output and monitor units was found to vary by as much as 10%. The central-axis depth dose was found to remain constant within experimental error. Field flatness measurements indicated that for lower power outputs, the radiation field was less flat than for higher power outputs. A mechanism for the change in calibration is suggested based upon field flatness measurements.