Analytical evaluation of methods for serum creatine kinase-MB. Electrophoresis, immunoinhibition and solid phase
R H Gadsden1, C N Papadea, J C Cate
1Department of Pathology and Laboratory Medicine, Medical University of South Carolina, Charleston 29425.
Abstract:
The purpose of this study was to evaluate immunoassay methods for the measurement of serum cardiac creatine kinase isoenzyme (CK-MB) with respect to sensitivity and specificity. The CK-MB electrophoretic assay (Helena Laboratories) was used as the reference. Two principles of immunoassay were included in the evaluation,--immunoinhibition and solid phase separation. The direct immunoinhibition techniques were from Beckman Instruments (CKMB reagent) and DuPont Medical Products (CKMB). Three solid phase separation techniques were from Abbott Laboratories (IMx CKMB), DuPont (acaPlus MCKMB), and Tosoh Medics Inc. (AIA-Pack CKMB). The electrophoretic method for separation of the CK isoenzymes has good specificity but lacks sensitivity for CK-MB in low concentrations. The immunoinhibition methods lack specificity and correlate poorly with the electrophoresis method and with the solid phase methods. The solid phase separation techniques are highly sensitive and show an excellent correlation with electrophoresis when based on specificity. The solid phase separation methods correlate well with each other.
Related Concept Videos
Two-dimensional Gel Electrophoresis
The first dimension separation uses the isoelectric focusing or IEF technique performed on immobilized pH gradient (IPG) strips that separate proteins according to their isoelectric points.
Biological samples, such as cells...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.