External cardiac pacing using low impedance electrodes suitable for defibrillation: a comparative blinded study

R H Falk1, N J Battinelli

  • 1Department of Medicine, Boston City Hospital, Massachusetts 02118.

Journal of the American College of Cardiology
|November 1, 1993
PubMed
Summary

High impedance electrodes do not improve external cardiac pacing outcomes. Low impedance electrodes, suitable for defibrillation, are effective for pacing, simplifying procedures with a single electrode set.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Calcium-Proton Exchange During Algal Calcification.

The Biological bulletin·2018
Same author

Amyloid diseases of the heart: assessment, diagnosis, and referral.

Heart (British Cardiac Society)·2010
Same author

High-dose intravenous melphalan and autologous stem cell transplantation as initial therapy or following two cycles of oral chemotherapy for the treatment of AL amyloidosis: results of a prospective randomized trial.

Bone marrow transplantation·2003
Same author

An overview of the use of high-dose melphalan with autologous stem cell transplantation for the treatment of AL amyloidosis.

Bone marrow transplantation·2001
Same author

Amyloidosis.

Postgraduate medical journal·2001
Same author

Infusion of light chains from patients with cardiac amyloidosis causes diastolic dysfunction in isolated mouse hearts.

Circulation·2001

Area of Science:

  • Cardiology
  • Biomedical Engineering
  • Medical Devices

Background:

  • External cardiac pacing traditionally uses high impedance electrodes (> 500 Ω).
  • High impedance electrodes are unsuitable for defibrillation, which requires low impedance.
  • A need exists for a single electrode set for both pacing and defibrillation.

Purpose of the Study:

  • To investigate if electrode impedance affects cardiac pacing thresholds.
  • To determine the feasibility of using low impedance electrodes for external pacing.

Main Methods:

  • 110 external cardiac pacing attempts in 32 subjects.
  • Used high or low impedance electrodes with commercial pacemakers.
  • Randomized double-blind design to assess pacing thresholds and discomfort.

Related Experiment Videos

Main Results:

  • No significant difference in pacing thresholds or discomfort across electrode types and impulse durations.
  • Mean pacing thresholds ranged from 72.5 to 77.5 mA.
  • Individual pacing thresholds varied widely among subjects.

Conclusions:

  • High impedance electrodes offer no advantage for external pacing.
  • External pacing is feasible with low impedance electrodes suitable for defibrillation.
  • A single external pacing-defibrillation electrode can be used for both procedures.