Comparative performance of five commercial prick skin test devices
H S Nelson1, D M Rosloniec, L I McCall
1Department of Medicine, National Jewish Center for Immunology and Respiratory Medicine, Denver, CO 80206.
The Journal of Allergy and Clinical Immunology
|November 1, 1993
Summary
Different skin prick test devices vary in skin trauma, patient acceptance, and accuracy. Device choice impacts allergen testing interpretation, requiring specific wheal sizes for reliable results.
Area of Science:
- Allergy and Immunology
- Dermatology
- Medical Device Evaluation
Background:
- Skin prick testing is a common diagnostic tool for allergies.
- Device selection can influence test reproducibility and interpretation.
- Standardization of skin prick testing is crucial for accurate allergy diagnosis.
Purpose of the Study:
- To compare the performance of five commercial skin prick test devices.
- To evaluate device reproducibility, patient acceptance, and false-negative rates.
- To determine device-specific wheal sizes for reliable allergen testing.
Main Methods:
- Comparison of five skin prick test devices: Hollister-Stier, ALK Laboratories, bifurcated needle (prick/puncture), Multi-Test, and DermaPIK.
- Assessment of reproducibility using histamine base (10 mg/ml).
- Evaluation of skin trauma, patient acceptance, and reaction sizes at control sites.
Main Results:
- All devices showed similar reproducibility with histamine.
- Skin trauma varied significantly, with Multi-Test and DermaPIK being most traumatic.
- More traumatic devices led to larger control site reactions and lower patient acceptance.
- Different minimum wheal sizes are proposed for 99% specificity for each device.
Conclusions:
- Device-induced skin trauma influences skin prick test results and interpretation.
- Specific wheal size thresholds are recommended for each device to ensure diagnostic accuracy.
- Allergen testing interpretation must account for device-specific characteristics and potential for false negatives.