Quality audit of megavoltage radiotherapy units: intercomparison of dose at a reference point using a mailed
1Radiation Oncology Department, University Hospital, Zurich, Switzerland.
Abstract:
A dosimetry intercomparison based on mailed TL-dosimeters and the well proven IAEA/EORTC phantom is described. Its aim is to identify discrepancies in dosimetry larger than +/- 3%. Dosimeters were mailed to all radiotherapy centres in Switzerland for irradiation with 2 Gy at a reference point in a water container, using photons and electrons. Thirty-six beams were monitored. The results show an agreement of within 2% for the majority of the beams monitored. Two electron beams were at 6% of the reference value.