Evaluation of a new vascular occlusion device
G K Nazarian1, Z Qian, Z Vlodaver
1Department of Radiology, University of Minnesota Hospital and Clinic, Minneapolis 55455-0392.
Investigative Radiology
|December 1, 1993
Summary
A new vascular plug device effectively occluded canine arteries and patent ductus arteriosus (PDA) in animal models. While promising, further modifications are needed to minimize misplacement risks.
Area of Science:
- Vascular Surgery
- Interventional Cardiology
- Biomedical Engineering
Background:
- Vascular occlusion devices are crucial for treating various cardiovascular conditions.
- Existing devices like coils have limitations, necessitating development of novel solutions.
- The patent ductus arteriosus (PDA) remains a target for innovative occlusion techniques.
Purpose of the Study:
- To evaluate a novel balloon-expandable vascular plug for efficacy in canine femoral/iliac arteries.
- To assess the device's performance in a canine model of patent ductus arteriosus (PDA).
- To determine the safety and potential limitations of the new vascular occlusion device.
Main Methods:
- Four vascular plug variations were implanted in 18 canine iliac/femoral arteries.
- Angiography was used for follow-up at 1-2 hours post-implantation.
- The 'butterfly' plug was tested in a canine PDA model, with long-term follow-up in some animals.
Main Results:
- Complete or near-complete occlusion was achieved in 12 of 18 arteries within 2 hours.
- Successful PDA occlusion occurred in 6 of 8 dogs within 1-2 days.
- One instance of vessel wall erosion occurred at 3 months without sequelae; three misplacements led to one fatality.
Conclusions:
- The balloon-expandable vascular plug shows promise for vascular occlusion, with no observed migration.
- Device sizing may be less critical compared to coils due to its design.
- Delivery system modifications and increased experience are required to mitigate misplacement risks.