Repeatability of a specially designed intraoral colorimeter
1Division of Restorative Dentistry and Prosthodontic Sciences, New York University, College of Dentistry, New York.
The Journal of Prosthetic Dentistry
|June 1, 1993
Summary
This study found that a modified intraoral colorimeter showed poor repeatability for measuring denture tooth color over time, even after recalibration. This indicates potential challenges in accurately assessing shade matching for dental prosthetics.
Area of Science:
- Dental Materials Science
- Colorimetry
- Prosthodontics
Background:
- Accurate color measurement is crucial for dental prosthetics.
- Existing colorimeters may require modification for intraoral use.
- Assessing the long-term repeatability of intraoral colorimeters is essential for clinical reliability.
Purpose of the Study:
- To evaluate the repeatability of a modified Minolta CS-100 colorimeter with a DP-101 data processor for intraoral color measurements.
- To assess color stability of porcelain and acrylic resin denture teeth.
- To determine the clinical applicability of the device for shade matching.
Main Methods:
- A modified Minolta CS-100 colorimeter was used for intraoral measurements.
- Porcelain and acrylic resin denture teeth (n=5 each) were tested.
- L*a*b* color readings were taken on the same day, day 3, and day 8, with and without recalibration.
Main Results:
- Same-day repeatability showed delta E values < 0.4 for most samples.
- Repeatability decreased significantly by day 3 (delta E < 0.4 in only 2/10 samples).
- Day 8 results after recalibration showed delta E values > 0.4 in all samples (0.57–2.75), with in vivo testing yielding delta E values from 1.1 to 32.1.
Conclusions:
- The modified intraoral colorimeter demonstrated poor long-term repeatability for denture tooth color measurements.
- Recalibration did not consistently restore acceptable color measurement accuracy.
- The findings suggest limitations in using this device for reliable clinical shade matching of dental prosthetics over time.