A biomechanical evaluation of an intramedullary fixation device for intertrochanteric fractures
C M Flahiff1, C L Nelson, J M Gruenwald
1Department of Orthopaedic Surgery, University of Arkansas for Medical Sciences, Little Rock 72205.
Abstract:
The strength and stability of an intramedullary device when used to fix intertrochanteric fractures were determined and compared with the dynamic hip screw (DHS). A standard four-part osteotomy was created in eight paired fresh frozen human cadaver femurs. The intramedullary fixation device and a DHS were implanted in each pair member, and mechanical testing was performed. Micromotion was measured during cyclic loading to determine implant stability, and then the specimens were loaded to failure. The intramedullary fixation device had significantly greater stability in cyclic loading than the DHS and required more than twice the force for failure. For patients with osteoporosis, this device may be a useful alternative to standard sliding nail systems.