Related Experiment Videos
Solid phantom material for the dosimetry of iodine-125 seed ophthalmic plaques
K P Hermann1, W Alberti, P Tabor
1Institut für Medizinische Physik und Biophysik, Universität Göttingen, Germany.
Purpose:
A tissue-equivalent solid phantom material, RE-1, closely simulating the radiological attenuation and scattering properties of the human eye for the iodine-125 photon spectrum and their Compton-scattered secondary photons, was fabricated on a polyethylene base with CaCO3 and MgO as inorganic additives.
Methods And Materials:
A 24 mm diameter spherical phantom was made from 1.1 mm thick sheets of RE-1, and holes were drilled in which 1 mm3 TLD cubes were placed.
Results:
The radial dose function g(r), which determines the dose profile on the transverse axis, was measured in a quasi-infinite phantom of RE-1.
Conclusion:
The values obtained deviate only slightly from those for a quasi-infinite phantom made from water-equivalent material.