Criteria to select proper valve prosthesis for aortic valve replacement. Comparative assessment of various valve

M Shigenobu1, S Sano

  • 1Department of Cardiovascular Surgery, Okayama University Medical School, Japan.

Insights

This study compared pressure gradients across four aortic valve prostheses. St. Jude Medical and Björk-Shiley valves showed better performance, while Lillehei-Kaster and Omniscience valves had higher gradients, necessitating careful patient follow-up.

Area of Science:

  • Cardiovascular Surgery
  • Biomedical Engineering
  • Clinical Cardiology

Background:

  • Aortic valve prostheses are crucial for managing aortic valve disease.
  • Different prosthetic valve designs may impact hemodynamic performance.
  • Understanding pressure gradients is key to assessing valve function.

Purpose of the Study:

  • To compare pressure gradients (PG) across four types of aortic valve prostheses: St. Jude Medical (SJM), Björk-Shiley (B-S), Lillehei-Kaster (L-K), and Omniscience (O-S).
  • To evaluate the relationship between valve area index (VAI) and PG for each prosthesis type.
  • To identify optimal valve selection for specific patient body surface areas to maintain low pressure gradients.

Main Methods:

  • Continuous wave Doppler echocardiography was employed to measure pressure gradients in 147 patients across four prosthetic valves.
  • Cineradiography was used to determine disc-opening angles for the L-K and O-S valves.
  • Statistical analysis, including correlation studies between VAI and PG, was performed.

Main Results:

  • St. Jude Medical valves exhibited consistently low pressure gradients.
  • A significant negative correlation between VAI and PG was observed for SJM and B-S valves.
  • Lillehei-Kaster and Omniscience valves showed no significant VAI-PG correlation, with higher instances of elevated pressure gradients (>50 mmHg) due to suboptimal disc-opening angles.

Conclusions:

  • St. Jude Medical and Björk-Shiley valves demonstrate favorable hemodynamic profiles, with specific sizes recommended for patients with body surface areas of 1.3-1.7 m2 to maintain PG below 20 mmHg.
  • The suboptimal disc-opening angles of Lillehei-Kaster and Omniscience valves contribute to higher pressure gradients.
  • Patients with L-K and O-S prostheses require vigilant monitoring due to the increased risk of significant pressure gradients.

Related Concept Videos

Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Mitral Valve Prolapse II: Assessment and Management01:22

Mitral Valve Prolapse II: Assessment and Management

IntroductionA range of clinical features characterizes Mitral Valve Prolapse (MVP), but it is important to note that many individuals with MVP are asymptomatic and may remain so throughout their lives. For those who do exhibit symptoms, the following are the key clinical features:Palpitations: This is a common symptom where individuals feel an irregular or rapid heartbeat. Palpitations in MVP are often due to arrhythmias such as premature ventricular contractions or supraventricular tachycardia.
Aortic Regurgitation II: Clinical Features and Diagnostic Tests01:22

Aortic Regurgitation II: Clinical Features and Diagnostic Tests

Aortic valve regurgitation (AR) occurs when the aortic valve fails to close properly, allowing blood to flow backward from the aorta into the left ventricle. This backflow can result in two distinct clinical presentations: acute and chronic AR, each characterized by its own set of symptoms and physical findings.Acute Aortic RegurgitationAcute AR presents with a sudden onset of severe symptoms. Patients typically experience profound dyspnea (shortness of breath), chest pain, and signs of left...