Failure of a well-fixed bone-ingrown titanium hip prosthesis

W A Hodge1, J P Collier, V A Surprenant

  • 1Department of Orthopaedic Surgery, Good Samaritan Medical Center, West Palm Beach, Florida.

Orthopaedic Review
|June 1, 1993
PubMed
Summary

Revision surgery for a cementless titanium femoral stem revealed significant titanium and polyethylene wear debris. Detached fiber-mesh pads from the stem were identified as a debris source, contributing to osteolysis.