Comparative study of the mechanical behaviour of Castaman external fixator in different mounting configurations
E Guglielmino1, G La Rosa, W Leonardi
1Institute of Civil Engineering and Energetics, University of Reggio Calabria, Italy.
Abstract:
The mechanical behaviour of the Castaman external fixator was determined using different mounting configurations in a system that simulated a long bone fracture during the formation of the bone callus. The first stage of the study was the determination of the stress-strain characteristics of the system with monolateral mounting. The second stage, the object of this study, is the analysis of the system with bilateral mounting and a comparison between the two configurations of the external fixator examined. Strain gauge analysis was used to determine the strains and stresses in the various elements of the system (fiches and external unit). Numerical techniques (FEM) were also used in order to confirm the results obtained using the simulation system. The results obtained show clearly the greater stiffness of the bilateral system. Combined bending and compressive stresses are present on both the screws and the external units, and reveal a little asymmetrical behaviour due to the conicity of the screws.
More Related Videos
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...
Stability of structures