Implementation and optimization by the simplex method of a 3D double echo sequence in steady-state free precession
I Dufour1, J Bittoun, I Idy-Peretti
1C.I.E.R.M., Hôpital de Bicêtre, Le Kremlin, France.
Abstract:
Under steady-state conditions, the resulting echoes have very complex T1 and T2 relationships. Many authors exploited these echoes in different sequences to produce either T1- or T2-weighted images. The simultaneous acquisition of two echoes in a single sequence provides two images of clearly different contrasts. We implemented such a sequence, in a 3D-acquisition mode, combining the advantages of thin and contiguous slices with those of a multi-echo sequence. The contrast of the images was correlated with theoretical results, derived from Bloch equations. In order to estimate the acquisition parameters (alpha, TE, TR) to obtain an optimal T1- or T2-contrast between two tissues, a computer simulation of these equations was used in conjunction with the simplex method. The results show that this sequence improves the clinical efficiency of MRI, particularly in neurological and articular disease.
Related Concept Videos
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
NMR Spectrometers: Resolution and Error Correction
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...