Pull-out strength of five suture anchors
J E Carpenter1, D N Fish, L J Huston
1Orthopaedic Research Laboratories, University of Michigan, Ann Arbor.
Summary
Suture anchors performed best in thicker bone, away from the joint. The Mitek GII anchor showed the lowest failure rate in pull-out tests, indicating superior fixation strength for orthopedic procedures.
Area of Science:
- Orthopedic biomechanics
- Biomaterials science
- Surgical device engineering
Background:
- Suture anchors are crucial for soft tissue reattachment in orthopedic surgery.
- Understanding the biomechanical performance of different suture anchor designs is essential for optimizing surgical outcomes.
- Cadaveric models provide a valuable platform for evaluating device failure modes under controlled conditions.
Purpose of the Study:
- To compare the biomechanical properties and failure modes of five commercially available suture anchor devices.
- To identify factors influencing suture anchor performance, including bone quality and loading direction.
- To determine which suture anchor exhibits superior fixation strength in human cadaveric proximal tibiae.
Main Methods:
- Testing five suture anchor devices to failure in 198 trials using human cadaveric proximal tibiae.
- Evaluating anchor performance based on insertion success, pull-out strength, and failure mechanisms.
- Analyzing the effect of bone thickness, proximity to the joint surface, and loading direction on anchor fixation.
Main Results:
- The Mitek GII anchor demonstrated the lowest pull-out failure rate (19%) compared to others (46%).
- Anchors performed significantly better in thicker cortical bone, further from the joint, and when loaded parallel to the bone surface (p < 0.05).
- In perpendicular loading, Mitek GII and Statak anchors exhibited the highest mean failure loads (82.5 N and 90.2 N, respectively).
- The Acufex Rod TAG device experienced insertion failures (23%) in thicker cortical bone.
Conclusions:
- Suture anchor fixation strength is significantly influenced by bone quality and loading conditions.
- The Mitek GII anchor appears to offer superior pull-out resistance, making it a potentially reliable option for tibial fixation.
- Careful consideration of anchor placement and anticipated loading is critical for successful soft tissue reattachment in orthopedic surgery.