Related Experiment Video
Updated: Aug 8, 2026

Solid Phase 11C-Methylation, Purification and Formulation for the Production of PET Tracers
Published on: October 24, 2019
Rapid radiochemical and chemical quality control of [11C]putrescine
S J Gatley1, R A Ferrieri, J S Fowler
1Chemistry Department, Brookhaven National Laboratory, Upton, NY 11973.
Abstract:
A short (4.6 x 50 mm) cation exchange column was used in conjunction with conductivity and radioactivity detectors to determine the radiochemical purity (> 99%) and specific radioactivity (0.5-1.0 Ci/mu mol) of [1-11C]putrescine prepared via Michael addition of [11C]cyanide to acrylonitrile. The absence of acrylonitrile, a rodent carcinogen, from the final preparation was verified at the 50 ng level capillary vapor-phase chromatography (VPC) using a nitrogen-phosphorus detector. Routine VPC analysis using a Poropak Q column and flame ionization detection showed that preparations contained no more than 1 microgram of acrylonitrile.
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
09:08Enhancing Efficiency and Radiolabeling Yields of Carbon-11 Radioligands for Clinical Research Using the Loop Method
Published on: December 20, 2024