A comparison of four glucose monitors in a hospital medical surgical setting
Abstract:
Four bedside glucose monitors were evaluated and compared for accuracy in correlation with concomitant laboratory glucose values. Of these four monitors--Diascan, Glucometer Q.A., Accucheck, and One Touch--two bedside glucose monitors were comparable with a correlation coefficient of 0.91 for Glucometer Q.A. and 0.90 for One Touch. These two machines were then evaluated for precision and linearity. Although the Glucometer Q.A. was more precise for all three levels of glucose, the One Touch linearity was almost perfect. Lastly, in asking nursing staff for an evaluation of One Touch and Glucometer Q.A., 15 of 16 staff members preferred One Touch.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Flow Sheet
Here's a closer look at the examples of flowsheets commonly used by nurses:
Graphic Sheet Documentation:
Holter Monitor: 24-Hour Monitoring