Related Experiment Videos
Comparison of half-band and full-band electrodes for intracochlear electrical stimulation
1Department of Otolaryngology, University of Melbourne, Australia.
Abstract:
It was hypothesized that intracochlear stimulating electrodes oriented toward the modiolus would require a lower stimulus current to elicit a threshold neural response than longitudinal band electrodes. Electrically evoked auditory brain stem responses (EABRs) and electrode impedances were recorded with full-band and oriented half-band scala tympani electrodes in anesthetized, deafened cats. To elicit a threshold EABR, the stimulus current required for stimulation through half-band electrodes oriented toward the modiolus was not significantly different from the current required for stimulation through full-band electrodes. The impedances of full-band electrodes were significantly lower than those of half-band electrodes. Considering the significantly higher impedance and current density of half-band electrodes in comparison to full-band electrodes, as well as the critical orientation of half-band electrodes during implantation, we believe that full-band electrodes have advantages over half-band electrodes for scala tympani implantation.