Oscillometric blood pressure measurements by different devices are not interchangeable
M A Kaufmann1, H Pargger, L J Drop
1Department of Anesthesia, Massachusetts General Hospital, Boston 02114, USA.
Anesthesia and Analgesia
|February 1, 1996
Summary
Noninvasive blood pressure (NIBP) monitors from different brands are not interchangeable. Studies show significant variability in readings between devices, exceeding established guidelines and impacting clinical decisions.
Area of Science:
- Medical Devices
- Clinical Measurement
- Cardiovascular Monitoring
Background:
- Noninvasive blood pressure (NIBP) monitoring is crucial in patient care.
- Oscillometric devices from different manufacturers are widely used.
- The interchangeability of NIBP readings across brands has not been consistently established.
Purpose of the Study:
- To evaluate the concordance of blood pressure (BP) measurements between different NIBP devices.
- To assess if readings from SpaceLabs (SpL) and Marquette (Marq) devices are interchangeable with Dinamap (DIN).
Main Methods:
- Simultaneous BP measurements were taken on opposite arms of 25 patients undergoing electroconvulsive therapy (ECT).
- Devices used included SpL, Marq, and Dinamap 1846SX (DIN).
- Bland-Altman plots and American Association for Advancement of Medical Instrumentation (AAMI) guidelines were used for agreement analysis.
Main Results:
- Dinamap devices showed high reproducibility (SDD: SBP 7 mm Hg, DBP 6.3 mm Hg).
- SpL vs. DIN showed higher variability (SDD: SBP 9.1 mm Hg, DBP 8.3 mm Hg) with significant mean differences.
- Marq vs. DIN also showed high variability (SDD: SBP 11.8 mm Hg, DBP 9.7 mm Hg).
- Systolic BP differences exceeding 10 mm Hg occurred in 31% (SpL vs. DIN) and 32% (Marq vs. DIN) of observations, surpassing AAMI guidelines.
Conclusions:
- Measurements from SpL and Marq NIBP devices are not interchangeable with Dinamap devices.
- Significant variability in BP readings exists between different oscillometric monitors.
- Clinical decisions should consider potential inaccuracies when using NIBP devices from different brands.