Related Experiment Videos
The evaluation of sizing accuracy of particle counters for parenteral drugs
O Fujishita1, T Sendo, A Hisazumi
1Department of Hospital Pharmacy, Faculty of Medicine, Kyushu University, Fukuoka, Japan.
Abstract:
Our evaluation of electronic liquid-borne particle counter systems has shown that, for accurate measurement of particulate matter in injections, the half count values of voltage thresholds from the particle counter itself should be used in the calibration for the particle sizes being evaluated. The manual method which uses the half count values in the USP XXIII<788> was improved, and the validity of our calibration method was supported by the results based on the ratio test, at 10 microns to those obtained at 15 microns is between 1.5 and 3.5, as per the Seventh Supplement of USP XXII <788>. Additionally, computer-based systems with automated calibration routines which solve the approximate Gaussian distribution of calibration particles have advantages. For the proper determination of the sizing accuracy of a particle counter, criteria of the ratio test at 10 microns and 15 microns as per the Seventh Supplement of USP XXII <788> should be applied at all critical particle sizes.