A small and lightweight three-channel signal-conditioning unit for strain-gage transducers: a technical note
J E Sanders1, L M Smith, F A Spelman
1Center for Bioengineering, Regional Primate Research Center, University of Washington, Seattle 98195, USA.
Abstract:
A small and lightweight signal conditioning unit for use with strain-gage transducers is described. The unit provides bridge-balancing, amplification, and filtering for three channels of Wheatstone bridge circuits. The electronics are housed with a 4.5 cm x 3.0 cm x 2.5 cm box that allows connection to a transducer and to a cable that extends to power supplies and a data storage facility. Evaluation tests showed the signal conditioning unit had low noise (0.356 microV, peak-to-peak, referenced to input) with crosstalk between channels of less than 0.02% of the peak-to-peak input signal. The signal conditioning unit has application in biomechanics where a small and lightweight unit is needed; for example, skin/support interface stress measurements in prosthetics and orthotics research.