Related Experiment Videos
The application of a modified mini-flexible-fiberoptic endoscopy in pediatric practice
1Children's Medical Center, Veterans General Hospital-Taipei, Taiwan, R.O.C.
Background:
The application of bronchoscopy in infants and small children is still restricted in most institutions because of their limited airways. The purpose of this study was to evaluate the usefulness of a modified mini (2.1 mm OD) flexible fiberoptic endoscope (FFE) in diagnosis and therapy of suspected pediatric airway problems.
Methods:
A prospective two-year study was carried out in a tertiary care neonatal/pediatric intensive care unit and general ward in a children medical center of a university-affiliated hospital. The employed mini-FFE (Machida ENT-30 F III, Tokyo, Japan) was modified with the addition of an external silastic catheter (0.5 mm inward diameter, 0.9 mm outward diameter) for suctioning, oxygen delivery, and medications during study.
Results:
Totally, 247 laryngoscopies and 212 bronchoscopies were performed in 207 patients without significant morbidity or any associated mortality. The age of patients ranged from 1 day to 10 years. About one-fifth of the procedures (larngoscopy 21.5%, 53/247; bronchoscopy 19.8%, 42/212) were performed in patients under the age of one month. Patient's body weight ranged from 650 g to 40 kg. About fifty-five percent of the procedures (laryngoscopy 55.1% 136/247; bronchoscopy 55.2%, 117/212) were performed in patients weighing less than 5 kg. The most common path of either laryngoscopy (94.3%, 233/247) or bronchoscopy (94.8%, 201/212) was through the nasal route. In 102 patients who had been demonstrated to have upper airway problems, 16 patients (15.7%) also proved to ahve significant lesion(s) below the glottis.
Conclusions:
This modified FFE is a safe and a valuable instrument that can serve as an important aid in the diagnosis and therapy as well as a guide for surgical intervention of respiratory tract disorders in small infants and children.