Evaluation of a rapid method of determination of plasma fibrinogen in swine
Related Experiment Video
Updated: Aug 10, 2026

Measurement of Factor V Activity in Human Plasma Using a Microplate Coagulation Assay
Published on: September 9, 2012
Abstract:
An evaluation was made of a rapid semiautomated method for determining fibrinogen level in swine plasma. This method, referred to as thrombin time method or fibrometer method, is based on the principle that when thrombin is added to suitably diluted plasma, the time of clotting is linearly related to the fibrinogen concentration. The linear regression model for the standard curve prepared using swine plasma had an r value of 0.998. A comparison between the fibrometer and the Grannis methods done on 189 swine plasma samples showed good correlation between these two mehtods (r value 0.847). It was concluded that although the fibrometer method may not be as precise as the Grannis method, it would still be acceptable for clinical use in swine.