Related Experiment Videos
[CAP system methods versus FAST methods in immunologic monitoring of specific immunotherapy in pollen allergy]
W Szymański1, S Chyrek-Borowska
1Kliniki Alergologii Akademii Medycznej w Białymstoku.
Abstract:
The aim was to study the value of the CAP system (Pharmacia) in comparison with the FAST method (Bio Whittaker) in determination of total IgE, specific IgE and IgG4 during specific immunotherapy (sIT). The studies were carried out on a group of 67 patients with pollinosis treated with various pollen allergen extracts. The immunological studies were performed before and after several courses of sIT. A higher reproducibility and sensitivity of the CAP System was shown. The values of specific IgE were somewhat higher on using the CAP system than using the FAST method. The CAP system seems to be of great value in diagnostics and monitoring sIT, especially in birch pollen allergy.