Related Experiment Video
Updated: Aug 11, 2026

Mechanical Ventilation Boot Camp Curriculum
Published on: March 12, 2018
A study of the incorrect use of ventilator disconnection alarms
R M Campbell1, A Sheikh, M M Crosse
1Department of Anaesthetics, Southampton General Hospital.
Abstract:
Pressure-sensitive ventilator disconnection alarms may fail to function correctly when used with discharging compliance ventilators such as the Manley Blease. The increase in gas flow on disconnection generates raised pressure owing to the resistance provided by components of the breathing system. Thirty-four anaesthetists, who were unaware of the nature of the survey, were observed to see if previous recommendations for the correct adjustment of the ventilator alarm were being followed. It was found that 77% of alarms were incorrectly adjusted and would not have alarmed in the event of a true disconnection occurring.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...