Safety evaluation of laparoscopically applied clips

T Papaioannou1, L Daykhovsky, W S Grundfest

  • 1Laser Research and Technology Development Program, Cedars-Sinai Medical Center, Los Angeles, California 90048, USA.

Journal of Laparoendoscopic Surgery
|April 1, 1996
PubMed
Summary

The Endo Clip II demonstrated superior security over the Ligaclip in laparoscopic surgery, requiring more force for clip dislodgment from tissues and tubing. Both devices showed no leakage during simulated blood circulation tests.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Nonparametric Analysis Of Time-Resolved Fluorescence Data Based On The Laguerre Expansion Technique.

Proceedings of the 25th Annual International Conference of the IEEE Engineering in Medicine and Biology Society : a new beginning for human health : 17-21 September, 2003, Cancun, Mexico. IEEE Engineering in Medicine and Biology Society...·2024
Same author

Biaxial sensing suture breakage warning system for robotic surgery.

Biomedical microdevices·2019
Same author

Endoscopic innovations.

Surgical technology international·2015
Same author

New laser and angioscopic technologies in medicine.

Surgical technology international·2015
Same author

Flexible designs for phase II comparative clinical trials involving two response variables.

Statistics in medicine·2014
Same author

Glucose and arterial blood pressure variability in Obstructive Sleep Apnea Syndrome.

European review for medical and pharmacological sciences·2013

Area of Science:

  • Surgical Devices and Technology
  • Biomedical Engineering
  • Minimally Invasive Surgery

Background:

  • Laparoscopic surgery relies on secure closure devices.
  • Evaluating the performance of surgical clips is crucial for patient safety.
  • Two common clip appliers, Endo Clip II and Ligaclip, require comparative analysis.

Purpose of the Study:

  • To compare the in vitro security of the Endo Clip II and Ligaclip appliers.
  • To assess clip performance regarding dislodgment resistance and leakage.
  • To determine the force required for dislodging clips applied to silicone tubing and vascular tissue.

Main Methods:

  • In vitro testing of laparoscopic clip appliers (Endo Clip II, Ligaclip).
  • Dislodgment tests performed transversely and at 45 degrees on silicone tubing and porcine vascular tissue.

Related Experiment Videos

  • Leakage tests conducted under simulated pulsatile blood circulation (approx. 800 mm Hg).
  • Measurement of maximum force required for clip dislodgment.
  • Main Results:

    • Endo Clip II required significantly more force for transverse and 45-degree dislodgment compared to Ligaclip on silicone tubing and vascular tissue.
    • Transverse dislodgment forces for Endo Clip II on tissue (556 +/- 146 g) were higher than Ligaclip (356 +/- 170 g).
    • No leakage was observed with either clip applier on silicone tubing or tissue during pulsatile circulation tests.

    Conclusions:

    • The Endo Clip II exhibits superior mechanical security compared to the Ligaclip based on dislodgment resistance.
    • Both laparoscopic clip appliers provide equivalent, effective sealing against leakage in simulated surgical conditions.
    • These findings support the selection of Endo Clip II for enhanced security in laparoscopic procedures where clip integrity is paramount.