Dentin bonding system. Part II: Effect of crosshead speed
1Materials Science and Engineering Laboratory, National Institute of Standards and Technology, Gaithersburg, MD, USA.
Bio-Medical Materials and Engineering
|January 1, 1996
Summary
Dentin bond strength testing reveals no strain rate sensitivity below 1.0 mm/min. Above this speed, tensile bond strength becomes dependent on the crosshead speed, indicating interface-related mechanical responses.
Area of Science:
- Dental Materials Science
- Biomaterials Engineering
- Mechanical Testing
Background:
- Dentin bond strengths are typically assessed using mechanical tests (tensile/shear) at varying crosshead speeds.
- Commonly used speeds range from 0.1 to 10.0 mm/min, but the rationale for selection is often absent in literature.
- Optimal testing conditions aim to induce fracture without plastic deformation.
Purpose of the Study:
- To investigate the influence of crosshead speed on dentin bond strengths.
- To determine the strain rate sensitivity of dentin bond strengths.
- To elucidate the mechanical behavior of the bonding interface during tensile testing.
Main Methods:
- Tensile bond strengths were evaluated using the Scotchbond Multiple-Purpose/Z100 bonding system.
- Testing was performed across a range of crosshead speeds: 0.1, 0.5, 1.0, 5.0, and 10.0 mm/min.
- Analysis focused on identifying strain rate dependency in bond strength measurements.
Main Results:
- No significant strain rate sensitivity was observed when testing below a crosshead speed of 1.0 mm/min.
- Tensile bond strengths exhibited a clear strain rate dependency when the crosshead speed exceeded 1.0 mm/min.
- The composite resin (Z100) itself did not show strain rate dependency.
Conclusions:
- The observed strain rate dependency in tensile bond strength is primarily attributed to the mechanical response of the bonding interface.
- Crosshead speed selection is critical for accurate dentin bond strength evaluation, particularly to avoid interface-dependent artifacts.
- Understanding strain rate effects is crucial for reliable assessment of dental adhesive performance.