The calibration of automated instruments for accuracy in hemoglobinometry
American Journal of Clinical Pathology
|July 1, 1977
Summary
College of American Pathologists surveys reveal manual hemoglobin tests show higher values than Coulter S measurements, averaging 0.3 g/dl difference. A new calibration method is recommended over commercial controls for accuracy.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Hematology
Background:
- Accurate hemoglobin determination is crucial for diagnosing and monitoring various medical conditions.
- Existing methods for hemoglobin measurement, including the Coulter S, may exhibit biases compared to reference standards.
- Ensuring reliable calibration is essential for the accuracy of automated hematology analyzers.
Purpose of the Study:
- To identify and quantify the bias between manual hemoglobin reference methods and Coulter S measurements.
- To evaluate the effectiveness of current calibration practices for the Coulter S analyzer.
- To recommend an optimal calibration and quality assurance strategy for hemoglobin determinations.
Main Methods:
- Analysis of survey data from the College of American Pathologists.
- Comparison of hemoglobin values obtained from manual reference methods and Coulter S automated analyzer.
- Evaluation of commonly used calibration techniques for the Coulter S system.
Main Results:
- A consistent bias was observed, with Coulter S hemoglobin values being lower than manual reference methods.
- The average difference in hemoglobin values was found to be 0.3 g/dl.
- Current calibration methods for the Coulter S analyzer appear to be suboptimal.
Conclusions:
- The proposed hemoglobin calibration and quality assurance method by Bull and colleagues is recommended as the most acceptable current approach.
- Commercial hemoglobin control materials are not suitable for calibrating hemoglobin measurements.
- Implementing improved calibration strategies is necessary to address the observed bias in Coulter S hemoglobin determinations.
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Counting is the type of measurement that is free from uncertainty, provided the number of objects being counted does not change during the process. Such measurements result in exact numbers. By counting the eggs in a carton, for instance, one can determine exactly how many eggs are there in the carton. Similarly, the numbers of defined quantities are also exact. For example, 1 foot is exactly 12 inches, 1 inch is exactly 2.54 centimeters, and 1 gram is exactly 0.001 kilograms. Quantities...
Guidelines For Measuring Vital Signs
Following these guidelines can help nurses accurately measure vital signs, assess changes in patient conditions, and provide timely treatment when necessary. Adhering closely to the guidelines ensures the accuracy and reliability of the results.
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of blood pressure in brachial artery(one-step method)
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
Calibration Curves: Linear Least Squares
A calibration curve is a plot of the instrument's response against a series of known concentrations of a substance. This curve is used to set the instrument response levels, using the substance and its concentrations as standards. Alternatively, or additionally, an equation is fitted to the calibration curve plot and subsequently used to calculate the unknown concentrations of other samples reliably.
For data that follow a straight line, the standard method for fitting is the linear...
For data that follow a straight line, the standard method for fitting is the linear...
Instrument Calibration
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...