Three-point repositioning of axes: three-dimensional alignment procedure for electron microscope tomography using three markers
Related Experiment Video
Updated: Nov 4, 2026

Sample Preparation by 3D-Correlative Focused Ion Beam Milling for High-Resolution Cryo-Electron Tomography
Published on: October 25, 2021
R Jonges1, E de Moor, P N Boon
1Department of Medical Physics and Informatics, University of Amsterdam, The Netherlands.
Abstract:
A description is given of a new procedure to align series of tilted graphs, made with an electron microscope, for computer tomographic purposes. The procedure uses the coordinates of three projected markers to calculate parameters needed for the reconstruction. To that end the procedure computes the direction of the tilt-axis, the translation and rotation parameters, the tilt-angle of every micrograph, and the spatial coordinates of the individual markers with their centre of gravity as origin of the coordinate system. A searching technique, based on cross-correlation, is described to locate accurately the micrographs markers.