Evaluation of a new pulse oximeter testing device

J A Fisher1, T Martire, G A Volgyesi

  • 1Department of Anaesthesia, Mount Sinai Hospital, Toronto, Ontario.

Canadian Journal of Anaesthesia = Journal Canadien D'Anesthesie
|February 1, 1996
PubMed
Summary

Routine testing of pulse oximeters is challenging. The Pulse Oximeter Tester (POT) provides a suitable stimulus for accurate pulse oximeter testing, demonstrating high specificity in identifying device performance.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Aqueous-phase mechanism for secondary organic aerosol formation from isoprene: application to the Southeast United States and co-benefit of SO<sub>2</sub> emission controls.

Atmospheric chemistry and physics·2020
Same author

Improved White Matter Cerebrovascular Reactivity after Revascularization in Patients with Steno-Occlusive Disease.

AJNR. American journal of neuroradiology·2018
Same author

Organic nitrate chemistry and its implications for nitrogen budgets in an isoprene- and monoterpene-rich atmosphere: constraints from aircraft (SEAC<sup>4</sup>RS) and ground-based (SOAS) observations in the Southeast US.

Atmospheric chemistry and physics·2018
Same author

Technology III: in-line vaporizer with reflector.

Journal of clinical monitoring and computing·2017
Same author

High volatile anaesthetic conservation with a digital in-line vaporizer and a reflector.

Acta anaesthesiologica Scandinavica·2017
Same author

Effects of acute controlled changes in end-tidal carbon dioxide on the diameter of the optic nerve sheath: a transorbital ultrasonographic study in healthy volunteers.

Anaesthesia·2017

Area of Science:

  • Biomedical engineering
  • Medical device testing
  • Clinical diagnostics

Background:

  • Routine testing of pulse oximeters and sensors presents significant challenges.
  • A valid testing device requires a pulsatile signal and tissue-like light absorption.
  • The Pulse Oximeter Tester (POT) is a novel device designed to address these testing limitations.

Purpose of the Study:

  • To evaluate the efficacy of the Pulse Oximeter Tester (POT) as a stimulus for pulse oximeter evaluation.
  • Assessing the POT's capability to provide a reliable testing environment for pulse oximetry devices.

Main Methods:

  • Pulse oximeters and sensors were tested using POTs simulating 80%, 90%, and 100% blood oxygen saturation.
  • Simulated heart rates of 30, 75, and 110 bpm were used during testing.

Related Experiment Videos

  • Performance was assessed by comparing readings from POTs to device outputs.
  • Main Results:

    • POTs yielded accurate SpO2 readings: 80.7 ± 1.3% for 80%, 90.3 ± 0.9% for 90%, and 100 ± 0.0% for 100% saturation.
    • No significant variations in readings were observed across different simulated heart rates.
    • Two pulse oximeters showed readings deviating by more than 2 SD, with subsequent identification of defective sensors.

    Conclusions:

    • Pulse Oximeter Testers (POTs) are suitable and effective stimuli for routine pulse oximeter testing.
    • The POT demonstrated high specificity in the tested sample.
    • The sensitivity of POTs for pulse oximeter testing remains to be determined.