Related Experiment Video
Updated: Aug 8, 2026

Glass Wool Filters for Concentrating Waterborne Viruses and Agricultural Zoonotic Pathogens
Published on: March 3, 2012
Bacterial challenge of NISSHO ultrafilter ETF 609: results of in vitro testing
S Krautzig1, G Lonnemann, S Shaldon
1Department of Nephrology, Medical School of Hannover, Germany.
Abstract:
In hemodialysis, a certain degree of bacterial contamination on the dialysate side is a regular finding. Concern has been growing that this contamination may lead to a chronic inflammatory response in the patient. Ultrafiltration of dialysate can be used to reduce bacterial content and levels of cytokine-inducing substances upstream of the patient's dialyzer. The aim of this study was to test in vitro the rejection capacity of a polysulfone hollow-fiber ultrafilter (ETF 609, NISSHO Co., Osaka, Japan) challenged with bacterial filtrates derived from Pseudomonas aeruginosa PA103. Results showed a reduction of interleukin-1 beta-inducing activity (measured on peripheral blood mononuclear cells) from 5,035 +/- 394 pg/ml prefilter to nondetectable levels postfilter and endotoxin levels (limulus amebocyte lysate assay) of 4,167 +/- 1,079 versus 12 +/- 2 pg/ml, respectively. In conclusion, ultrafiltration of dialysate with the polysulfone ultrafilter ETF 609 leads to a potent reduction of cytokine-inducing activity.