Periodontal probe precision using 4 different periodontal probes
L Mayfield1, G Bratthall, R Attström
1Department of Periodontology, Lund University, Malmö, Sweden.
Journal of Clinical Periodontology
|February 1, 1996
Summary
The manual periodontal probe demonstrated the lowest variation in measurements. Automated probes showed potential for deeper pocket depth readings, possibly due to examiner unfamiliarity.
Area of Science:
- Dentistry
- Periodontology
- Biomedical Engineering
Background:
- Periodontal probing is crucial for diagnosing and monitoring periodontal disease.
- Accurate and reproducible measurements are essential for effective treatment planning and evaluation.
- Various periodontal probes exist, differing in design and pressure application, potentially affecting measurement consistency.
Purpose of the Study:
- To compare the intra- and inter-examiner reproducibility of four distinct periodontal probes.
- To evaluate the influence of probe type (manual vs. electronic) and standardized pressure on measurement variability.
- To identify potential biases related to examiner experience with different probing technologies.
Main Methods:
- Four periodontal probes were tested: Hu-Friedy LL 20 (manual), Vivacare TPS (manual, 0.20 N), Vine Valley (electronic, 0.25 N), and Peri Probe Comp (computerized electronic, variable pressure).
- Two examiners performed duplicate probing measurements on 10 patients across 6 sites per tooth (molar, premolar, incisor).
- Measurements were repeated after one week, with systematic randomization of probe and examiner order to minimize bias.
Main Results:
- The manual Hu-Friedy LL 20 probe exhibited the lowest degree of variation (correlation coefficient of 0.83).
- The manual and Peri Probe Comp probes tended to record deeper probing pocket depths compared to the TPS and Vine Valley probes.
- A potential influence of examiner unfamiliarity with automated probes on the results was noted.
Conclusions:
- Manual periodontal probes may offer higher reproducibility compared to some electronic or computerized systems.
- Differences in recorded probing depths suggest probe design and pressure standardization significantly impact measurements.
- Further research with experienced users of automated probes is recommended to fully assess their reproducibility.