Radiation exposure to patients and personnel during interventional ERCP at a teaching institution
R L Heyd1, K K Kopecky, S Sherman
1Department of Radiology, Indiana University School of Medicine, Indiana University Medical Center, Indianapolis 46202-5253, USA.
Background:
This study was designed to determine the radiation dose to patients and personnel during ERCP procedures.
Methods:
Phantom studies were performed to compare dosimetry for two image capture systems and to determine the effectiveness of shielding in reducing stray radiation. Radiation techniques and dosimetry were recorded in 72 patients undergoing ERCP.
Results:
Phantom studies indicated that increasing fluoroscopy voltage from 75 to 96 kV decreased entrance dose by 50%. Image capture by digital radiography decreased radiation exposure by 66%. Shielding decreased stray radiation exposure by 93%. All patients underwent cholangiography (n = 71) and/or pancreatography (n = 53). The mean number of interventional ERCP procedures performed per patient was 1.8 (range, 0 to 6). The mean measured patient entrance dose was 80 mGy (8 R; range, 0.2 to 73 R); however, the calculated mean entrance dose (based on measured intensifying screen doses) may have been as high as 3000 mGy (30 R; range, 0.8 to 300 R). Measured patient exposure increased with fluoroscopy time (r = 0.9) and with the number of interventions performed (r = 0.3). The mean dose to personnel was estimated at 0.04 mR.
Conclusions:
The patient radiation dose depended most on fluoroscopy time. The dose may be lowered by minimizing fluoroscopy time, using higher voltage and lower current for fluoroscopy, and using digital radiography for documentation. Personnel were adequately protected.
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Endoscopic Procedures V: ERCP
Patient...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...