Theory of wear as related to the Björk-Shiley Delrin disc heart valve
1Mechanical Engineering Department, State University of New York at Binghamton 13902-6000, USA.
Background And Aims Of The Study:
Mechanical wear is an important consideration for the Björk-Shiley Delrin (BSD) heart valve, the disc of which is periodically impacted against the inlet strut by the momentum of the blood flow during closure. Impact wear theory was used in designing experiments as well as establishing theoretical evaluation and projections of wear life.
Materials And Methods:
The experimental apparatus involved a pivotal hammer device where the striking face could be varied by the inclusion of distinct spherical shapes; the sharper the radius, the higher the contact stress induced. The striking speed could be varied between one and three m/s, and the repetition rate between 2-50 Hz. Blood was simulated with an aqueous 42% glycerol solution. The same wear mechanism was induced on Delrin occluder disc specimens as the one governing the wear observed in Delrin occluder discs of some of the explanted BSD heart valves; thus two wear parameters, c and g, were established for the prediction of wear in BSD implants. The experimental runs were repeated on as many as four simultaneous specimens, for up to 7.5 x 10(7) cycles.
Results:
The wear history was found to comprise three distinct regions, corresponding to (a) initial plastic deformations; (b) zero wear; and (c) measurable wear. The zero wear region produced very little change in the contact dimensions, and represented a fatigue threshold to the progress of a wear scar. Procedures for analytical establishment of the zero wear limit are outlined in the paper. Measurable wear was governed by the process of increasing conformance between the repetitively contacting surfaces. The differential equation of wear was then used with the physical parameters c and g previously determined experimentally.
Conclusion:
The maximum depth of the wear scar after cycling the equivalent of 20 years was predicted to be 278 microns. This value appears to represent a safe dimension for BSD heart valves.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Bearing Stress
Due to the intricacy of these microforces, an average value, known as bearing stress, is often used by...
Plastic Deformation in Circular Shafts
Circular Shafts - Elastoplastic Materials
As torque on the...
Mitral Valve Prolapse I: Introduction
Design Example: Deciding Thickness of Lubricating Fluid in a Shaft
To calculate the required thickness of the lubricant layer, the tangential velocity at the shaft's surface must first be determined. This velocity is calculated by converting the rotational speed to angular velocity...