Related Experiment Videos
Reexamination of tympanic membrane temperature as a core temperature
1Marshall Dermatology Research Laboratories, Department of Dermatology, University of Iowa College of Medicine, Iowa City 52242, USA.
Journal of Applied Physiology (Bethesda, Md. : 1985)
|April 1, 1996
Summary
Tympanic temperature (Tty) can be a reliable core body temperature measure. A modified probe design minimizes facial cooling effects, making Tty a potentially standard method for core temperature determination.
Area of Science:
- Physiology
- Biomedical Engineering
Background:
- Tympanic temperature (Tty) measurement controversies exist regarding its accuracy for core body temperature.
- Previous studies show inconsistent effects of facial cooling/heating on Tty, limiting its clinical utility.
Purpose of the Study:
- To develop a safer, more reliable tympanic membrane probe.
- To investigate the causes of inconsistent Tty responses to facial thermal challenges.
- To validate Tty as an accurate measure of core body temperature.
Main Methods:
- Utilized a modified thermocouple probe with improved sensor design and insulation.
- Compared Tty measurements with esophageal temperature (Tes) under steady-state conditions.
- Assessed Tty and Tes responses to facial and scalp skin cooling and foot warming.
Main Results:
- Repositioning the probe sensor corrected Tty drops during facial cooling, indicating thermal contamination.
- A thermally insulated probe ring prevented Tty decreases during facial cooling, unlike a larger sensor area probe.
- Tty closely matched Tes in steady-state conditions (36.83°C vs. 36.87°C).
- Facial cooling minimally affected Tty and Tes.
- Tty detected temperature changes 5 minutes faster than Tes during foot warming.
Conclusions:
- A modified tympanic membrane probe design minimizes thermal contamination, enhancing reliability.
- Tympanic temperature is a valid and potentially superior measure of core body temperature compared to esophageal temperature.
- Further probe development could establish Tty as a standard for core body temperature monitoring.