Related Experiment Videos
Microdosimetry of monoenergetic neutrons
1Brookhaven National Laboratory, Upton, New York 11973, USA.
Abstract:
Tissue spheres 0.25, 0.5, 1.0, 2.0, 4.0 and 8.0 micron in diameter were simulated using a wall-less spherical counter filled with propane-based tissue-equivalent gas. Microdosimetric spectra corresponding to these site sizes were measured for five neutron energies (0.22, 0.44, 1.5, 6 and 14 MeV) and the related mean values /yF and /yD were calculated for several site sizes and neutron energies. An elaborate calibration technique combining soft X rays, an 55Fe photon source and a collimated 244Cm alpha-particle source was used throughout the measurements. The spectra and their mean values are compared with theoretically calculated values for ICRU standard tissue. The agreement between the calculated and the measured data is good in spite of a systematic discrepancy, which could be attributed, in part, to the difference in elemental composition between the tissue-equivalent gas and plastic used in the counter and the ICRU standard tissue used in the calculations.