Evaluation of the YSI 2300 glucose analyzer: algorithm-corrected results are accurate and specific
J R Astles1, F A Sedor, J G Toffaletti
1Department of Pathology, Duke University Medical Center, Durham, NC 27710, USA.
Clinical Biochemistry
|February 1, 1996
Summary
The YSI 2300 analyzer, using its algorithm-corrected whole blood glucose results, demonstrated commutability with plasma glucose measurements. This method showed reliable performance across varying hematocrit levels and was unaffected by common interfering substances.
Area of Science:
- Clinical chemistry
- Biomedical instrumentation
- Diagnostic assay development
Background:
- Accurate glucose measurement is crucial for diabetes management.
- Commutability between whole blood and plasma glucose assays is essential for consistent patient care.
- Glucose oxidase methods can be susceptible to interferences from various compounds and hematocrit variations.
Purpose of the Study:
- To evaluate the commutability of whole blood glucose measurements with plasma glucose values.
- To assess the performance of different glucose analyzers and hematocrit correction methods.
- To identify potential interferences in glucose oxidase-based methods.
Main Methods:
- Comparison of plasma glucose results from the Ektachem 700 analyzer with whole blood glucose results from the Nova Stat Profile 6 and YSI 2300 analyzers.
- Evaluation of two hematocrit (Hct) correction methods for the YSI 2300: multiplication by 1.11 (YSI1.11) and an algorithm-based correction (YSIcor).
- Testing for interference from various chemical compounds on the analyzers.
Main Results:
- The YSI algorithm-corrected (YSIcor) whole blood glucose results showed 88% agreement within ±5% of plasma results.
- YSIcor results were unaffected by hematocrit variations and abnormal erythrocyte size.
- The Nova analyzer showed falsely decreased results at low Hct, while YSI1.11 showed falsely elevated results. Ascorbic acid interfered with Ektachem, and acetaminophen interfered with Nova.
Conclusions:
- The YSI algorithm-corrected whole blood glucose measurements are commutable with plasma glucose results.
- The YSI 2300 analyzer, with its algorithm-based hematocrit correction, demonstrated robustness against tested interferences, including acetaminophen and acetylsalicylic acid.
- This validated method ensures reliable glucose monitoring across different sample conditions.