Related Experiment Video
Updated: Aug 17, 2026

Measuring Carbon-based Contaminant Mineralization Using Combined CO2 Flux and Radiocarbon Analyses
Published on: October 21, 2016
Carbon monoxide within circle systems
N Harrison1, A C Knowles, E A Welchew
1Northern General Hospital NHS Trust, Sheffield.
Abstract:
After reports of carbon monoxide toxicity from the United States, the Medicines Control Agency issued a warning which recommended that soda lime used in circle breathing systems should not be allowed to dry out. We have measured carbon monoxide levels within a circle system in vitro (under a variety of conditions) and in vivo. Carbon monoxide was detected only when a patient was connected to the circle system. As levels in the expiratory limb of the circle system were significantly higher than those in the inspiratory limb, we conclude that the patient acts as the source of carbon monoxide.
More Related Videos
Related Concept Videos
The Carbon Cycle
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
Carbon Dioxide Transport in the Blood
Forms of CO2 Transport
1. Dissolved in plasma: A small percentage (7-10%) of CO2 is transported and dissolved directly in the plasma.
2. Carbaminohemoglobin: Just over 20% of CO2 is chemically bound to...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Single Pipe Systems
In a Type I problem, fluid properties (density and viscosity), pipe characteristics (including diameter, length, and surface roughness), and the flow rate or average velocity are known. The...
Carbon-dioxide Fixation