Related Experiment Videos
Evolution of atrial signals from a single lead VDD pacemaker. AddVent European Multicentre Study Group
1Bristol Heart Institute, United Kingdom.
Abstract:
The atrial sensing capabilities of a new single pass lead VDD pacing system (Pacesetter AddVent) were assessed in a prospective multicenter study of 101 implants during the period July 1994 through March 1996. The pacing lead (Pacesetter AV Plus) has a unique quadripolar 4-in-line connector and uses a pair of ring electrodes with an interelectrode spacing of 12 mm for atrial sensing. The mean age of the patients (51 men) was 73 years (range 19-91). Seventy-five patients had complete heart block; the others had 2:1 AV block. Wide variations were found in signal amplitude: mean P wave amplitude, measured over four cycles in the supine position, was 2.4 +/- 1.9 mV at implant, dropping to 1.9 +/- 1.7 mV predischarge, and remaining constant at follow-up but with a narrower range. Holter monitoring was undertaken in 24 patients, with a total of 550 monitored hours. Mean AV synchrony was 98.2% +/- 4.6% (excluding premature ventricular contractions), with 20 patients (83%) showing > 99% AV synchrony, with atrial sensing at 0.1 mV where needed. No oversensing was observed in any patient. There was a low incidence of atrial fibrillation (2%) and sinus bradycardia (0%). The findings show that the range of atrial signals, although wide initially, converges over the first year and remains adequate for reliable AV synchronous pacing.