Ceramic surgical instruments: ex vivo evaluation of compatibility with MR imaging at 1.5 T
1Radnet and Future Diagnostics, Inc., Los Angeles, CA 90045, USA.
Journal of Magnetic Resonance Imaging : JMRI
|November 1, 1996
Abstract:
The purpose of this study was to evaluate the MR compatibility of ceramic instruments. Nine different ceramic instruments were tested with respect to ferromagnetism, heating, and artifacts using previously described techniques. There was no magnetic field attraction, temperature increases were < or = 1 degree C, and artifacts involved a signal void similar to the size and shape of each instrument. The ceramic instruments were demonstrated to be compatible with MR imaging and acceptable for use during intraprocedural MR imaging.