Related Experiment Video
Updated: Oct 2, 2026

Novel Diagnostics in Revision Arthroplasty: Implant Sonication and Multiplex Polymerase Chain Reaction
Published on: December 3, 2017
Separation and characterization of polyethylene wear debris from synovial fluid and tissue samples of revised knee
D L Wolfarth1, D W Han, G Bushar
1FDA Center for Devices and Radiological Health, Rockville, Maryland 20857, USA.
Abstract:
A study was made of in vivo-generated polyethylene wear particles as separated from synovial fluid samples and from tissue samples surrounding total knee arthroplasty. A comparison of particle size and morphology between the two particle groups was made to assess any effects of selective tissue capture, and macrophage encapsulation and digestion. In addition, a Raman spectroscopy technique was evaluated that enables positive identification of individual wear particles. The particles of the same size range found in the synovial fluid and tissue samples exhibited a comparable morphology. Notably, submicron-sized debris was present in both the synovial fluid and tissue samples surrounding knees with osteolysis. The novel micro-Raman analysis of individual particles was successful in the categorizing of wear debris as polyethylene or nonpolyethylene.

